The Lasair III Particle Counter is available in three different flow rate configurations (1 CFM, 50 LPM and 100 LPM) to meet a variety of cleanroom applications. It can be used for routine remote and mobile cleanroom monitoring. Interactive software steps the user through the ISO 14644-1 (including the latest 2015 revision), EC GMP Annex 1. 30 may 2015), iNStruCtioN maNualS, lexmark priNter, Smart 1500 upS JouAN c4i MoBile ceNTriFuge, Chamber Dia 450mm, max SpeeD 7800 rpm,max oal D 3.6kg at 4000rpm, 230. LASAir iii 310c pArTicle MeASuriNg SySTeM rieTSchle Vc 50 MoToriSeD VAcuuM puMp lASAir iii 310A pArTicle MeASuriNg SySTeM geoTech g100 co2 rANge ANAlySer. The Lasair II particle counter provides complete features in a compact, convenient particle counter. Its great portability, ease of use, and chemically-resistant easy-to-clean cover have made it the particle monitoring standard for pharmaceutical and high tech industries. The Lasair II 350L and 550L models features a 50 liter/minute sampling. Manual Lasair II-Contador de Particulas. Click the start the download. DOWNLOAD PDF. Report this file. Description Download Manual Lasair II-Contador de Particulas Free in pdf format. Sponsored Ads. Account 40.77.167.37. About Us We believe everything in the internet must be free. So this tool was designed for. 3 If your shipment is undamaged, store the packing materials. It is important to store packing materials, because warranties may not apply if return shipping containers are inadequate. Packing Lists The HandiLaz 104000-1 is shipped with the following standard items.
Advertising
Without measurement there is no control.
BENEFITS
• ISO 21501-4 compliant with a built-in Pulse
Height Analyzer (PHA)
Modded games for windows pc. • 100 LPM, 50 LPM, or 1 CFM flow rates
available
– 100 LPM flow samples 1 m
3
in 10 minutes
• Large IR touchscreen for simplified use
• Quiet pump improves work environment
• Calculates and creates reports for ISO
• Quiet pump improves work environment
• Calculates and creates reports for ISO
14644-1, EC GMP Annex 1, China GMP, and
FS 209E
• 21 CFR Part 11 compliant download
• Comprehensive validation manual makes
• Comprehensive validation manual makes
it easy to meet pharmaceutical regulatory
requirements
Lasair Iii User Manual Free
• Operates in cleanrooms via a web browser:
Set up, sample, display, print, download
data
• Choose from ten languages for display and
printout
• Removable batteries with optional external Nba live 19 review.
charger for continuous mobile use
• Stores sampling recipes to reduce operator
errors
• Long-term data archiving with DataAnalyst
software
Usb driver 2.0 free download. My main computer was having a very annoying problem that prevented me from connecting any USB devices to it. It always prompted a message saying the USB driver cannot be installed. And when I went to the Device Manager, I saw the USB 2.0 Hub driver showing as. Download the latest drivers, firmware, and software for your USB 2.0 4-Port Hub.This is HP’s official website that will help automatically detect and download the correct drivers free of cost for your HP Computing and Printing products for Windows and Mac operating system.
• Local printout for immediate hardcopy
results
• USB electronic printouts for long term
storage
• CSV file download for easy data integration
• Web browser for simplified remote
• Web browser for simplified remote
operation
APPLICATIONS
• Cleanroom monitoring
• Facility certification for ISO, EC GMP, China
• Facility certification for ISO, EC GMP, China
GMP, FS 209E
• Trend analysis
• Statistical process control
• Troubleshooting particle excursions
• Manifold compatible
• Portable or dedicated use
• Statistical process control
• Troubleshooting particle excursions
• Manifold compatible
• Portable or dedicated use
Lasair 3
The Lasair III Particle Counter sets the standard for portable aerosol particle
counters and meets the requirements of ISO 21501-4.
The Lasair III Particle Counter is available in three different flow rate
configurations (1 CFM, 50 LPM and 100 LPM) to meet a variety of cleanroom
applications. It can be used for routine fixed and portable cleanroom
monitoring. Interactive software steps the user through the ISO 14644-1,
EC GMP Annex 1, China GMP, and FS 209E room certification process.
The results of cleanroom certifications are available either through a local
printout, downloadable via USB in a secure format, or through the use of
external software packages such as DataAnalyst, Facility Net software or
Pharmaceutical Net software.
The user interface is facilitated by the use of a large easy-to-read IR
touchscreen. The touchscreen eliminates interaction problems commonly
associated with resistive or capacitive touchscreens such as requiring the use
of a stylus or difficulties using a touchscreen while wearing gloves.
The rugged chemical-resistant enclosure is lightweight for portability and
has a streamlined design to minimize particle traps, making the unit easy
to clean. Power is provided to the unit through the use of the on-board
hot-swappable batteries. Power can also be provided through the use of an
external AC power source, which can simultaneously power the unit as well
as charge the on-board batteries.
The Lasair III Particle Counter can accept inputs from a wide range of
environmental sensors through the four built-in 4-20 mA inputs. Wireless
network communications is made easy with an externally-mounted wireless
adapter, simplifying communications with network systems. The unit can
be remotely managed via the built-in web browser function or through
Ethernet to Pharmaceutical Net , Facility Net, or FacilityPro™ software.
Lasair® III
Aerosol Particle Counter Models 310B, 310C, 350L, 5100
Advertising
Popular Brands
Popular manuals
Advertising
© 2012 Particle Measuring Systems, Inc. All rights reserved.
AUTHORIZED REPRESENTATIVE
specifications
081612
Particle Measuring Systems Headquarters
5475 Airport Blvd., Boulder, CO 80301, USA
Tel: +1 303 443-7100 +1-800-238-1801
FAX: +1 303 449-6870
Instrument Service & Support: +1 800-557-6363
Customer Response Center: +1 877-475-3317
5475 Airport Blvd., Boulder, CO 80301, USA
Tel: +1 303 443-7100 +1-800-238-1801
FAX: +1 303 449-6870
Instrument Service & Support: +1 800-557-6363
Customer Response Center: +1 877-475-3317
310B, 310C
350L
5100
Channels
310B: 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 µm
310C: 0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm
0.3, 0.5, 1.0, 5.0, 10.0, 25.0 µm 0.5, 1.0, 2.0, 5.0, 10.0, 25.0 µm
Flow rate
1.0 CFM (28.3 LPM) ± 5%
50 LPM ± 5%
100 LPM ± 5%
Calibration
Meets ISO 21501-4 requirements
Maximum concentration
a
> 1,380,270/ft
3
> 832,599/ft
3
> 688, 495/ft
3
Counting efficiency
50% ± 20% for most sensitive threshold. Meets ISO 21501-4 requirements.
Zero count
7.07 counts/m
3
(1CFM); 4.00 counts/m
3
(50 LPM); 2.00 counts/m
3
(100 LPM)
Data storage
3,000 complete data sets. 21 CFR 11 compliance. Long-term data storage/analysis with DataAnalyst.
Communication modes
Ethernet or RS-232; USB data downloading; optional wireless Ethernet
Controlling software
Facility Net, Pharmaceutical Net, FacilityPro, Microsoft® Internet Explorer® 5.0+, Firefox®
Remote operation
Remote web browser operation; real-time download to Facility Net or Pharmaceutical Net
Reports
USB reports; sample printouts; cleanroom certification reports for ISO, EC GMP, and FS 209E; averaging
Environmental sensors
Four 4-20 mA inputs
Languages
English, French, German, Italian, Japanese (Kanji), Korean, Mandarin Chinese (Traditional or
Simplified), Spanish, Russian
Display and printer
8.4” color VGA display (640 x 480); IR touch screen (IP65 rated); built-in thermal printer
External surface
Polycarbonate (PC)
Enclosure cleaning materials Bleach, ethyl/isopropyl alcohol, peroxide/quaternary ammonium solutions
Sample tubing ID
Sample tubing ID
b
3/8”, 8 m max. length
1/2”, 8 m max. length
3/4”, 8 m max. length
Sample output filtering
Internally filtered to > 99.97% at 0.3 µm
Power and battery
c
100 – 264 V, 50 – 60 Hz, 150 W
Lithium battery: estimated operation 3 hr (single) 6 hr (dual) for 1 CFM unit
Dimensions (h, w, d)
11.9 x 12.9 x 10.2 in (30.1 x 32.7 x 25.9 cm)
Weight
13.2 lb (6 kg) without battery, 16.5 lb (7.5 kg) with two optional batteries
Operating environment
Temperature: 32 – 86 °F (0 – 30 °C); Humidity: 5 – 95% RH non-condensing
All dimensions, weights and values not provided with a tolerance are +/-10% values.
a. Less than 10.0% coincidence loss at maximum recommended concentration.
b. For pharmaceutical applications, tubing length should equal 2m maximum.
c. Battery life is based on continuous operation running one (1) minute samples and printing every minute.
Lasair® is a registered trademark of Particle Measuring Systems, Inc.
FacilityPro™ is a trademark of Particle Measuring Systems, Inc.
All other trademarks are the property of their respective owners.
U.S. and foreign patents pending.
Particle Measuring Systems, Inc. reserves the right to change specifications without notice.
Lasair® III
Aerosol Particle Counter Models 310B, 310C, 350L, 5100
Advertising
Popular Brands
Popular manuals